Six samples were assayed in one FX kit run in replicates of 10 using a microplate. The samples were two controls (one normal and one prolonged), two patient samples with normal levels, and two patients with prolonged levels.
Precision Data
FX Concentration (%) | CV Within Run |
Normal (98 – 107%) | 1.9 – 12.0% |
Abnormal (23 – 32%) | 4.7 – 19.6% |
The central tendency for within-run precision of the DiaPharma Factor X kit is approximately 7 %CV.